Kipp & Zonen CSD1 Instruction Manual Download Page 24

 
 
 

 
 
 

24 

 

The sensitivity of FD1 is adjusted at the factory at 50 microvolt per Watt 
per square meter. This offers the opportunity to check the intensity of the 
beam. At 500 Watt per square meter, the output voltage should be 25 
millivolts. 

 
3.  Now take the test CSD1 that needs to be checked, and put it in the 

position of the reference.  

 
4.  Take 4 readings of D1 of the CSD1 (J18-1 and J18-2) at 90 degree 

intervals. The average is called FD1. 

 
5.  Put D2 of the test CSD1 in the position of D1. Take the readings of D2 of 

the test CSD1 (J18-2 and J18-5) in 2 opposite directions. The average is 
called FD2. 

 
6.  Put D3 of the test CSD1 in the position of D2. Take the readings of D3 of 

the test CSD1 (J18-2 and J18-6) in 2 opposite directions. The average is 
called FD3. 

 
7.  FD1 should equal REF1. If the deviation is more than 5% it is suggested 

to replace the resistors at location 1, RES1 to adjust the sensitivity. The 
value of the replacement resistors is a function of the resistors that are 
present, and the ratio REF1/FD1. First the resistors at location 1 are 
removed, and their total resistance is measured. After this the value of 
the replacement resistors REP1 can be calculated according to the 
following formula: 

 

REP1 = total old resistance * ( REF1 / FD1 ) 

 

Note:

 

T

he calibration resistance consists of two separate resistors.  

The total value can be calculated by adding the two resistor values.

 

 
8.  FD2 should equal 3.3 times REF1. If the deviation is more than 5% it is 

suggested to replace the resistors at location 2, RES2 to adjust the 
sensitivity. The value of the replacement resistors is a function of the 
resistors that are present, and the ratio REF1/FD2. 

Summary of Contents for CSD1

Page 1: ...INSTRUCTION MANUAL Kipp Zonen 0342 300 1 SUNSHINE DURATION METER...

Page 2: ......

Page 3: ...pp Zonen guarantees that the product delivered has been thoroughly tested to ensure that it meets its published specifications The warranty included in the conditions of delivery is valid only if the...

Page 4: ...4...

Page 5: ...e CSD 1 Name SUNSHINE DURATION SENSOR to which this declaration relates is in conformity with the following standards Imissions EN 50082 1 Group standard IEC 100 4 2 IEC 801 2 8 kV IEC 100 4 3 IEC 801...

Page 6: ...6...

Page 7: ...REMENT OF DIRECT RADIATION 15 1 6 USING THE HEATER 15 1 7 CALIBRATION QUALITY CONTROL 15 1 8 MAINTENANCE 16 1 9 OPTICAL THEORY 16 2 PROPERTIES OF CSD 1 19 2 1 SPECIFICATIONS OF THE SUNSHINE DURATION M...

Page 8: ...8...

Page 9: ...calculated from the differences in signal levels Great care has been taken to design the angular characteristics of the detector resulting in a measurement that can be used anywhere on Earth in a fixe...

Page 10: ...expand upon the above CSD 1 should be regularly maintained and if possible be re calibrated once every year in order to check the instrument stability These items are addressed in 1 7 and 1 8 More abo...

Page 11: ...e sensor needs power from a source that can provide 9 15 VDC at a power of 10 mW without heating on This can easily be supplied by a small solar panel or a battery When using the measurement of the di...

Page 12: ...ation location Installing like this the measurement is optimised When connecting the wiring make sure to shield the power and signal cables as much as possible Figure 1 3 1 installation of the CSD 1 T...

Page 13: ...13 Figure 1 3 2 a global view of the installation of the CSD 1...

Page 14: ...puts sunshine duration direct radiation Inputs heater low heater high power supply 1 4 MEASUREMENT OF SUNSHINE DURATION For measurement of sunshine duration connect the power supply input and the suns...

Page 15: ...ting the instrument CSD 1 has on board heaters that can be used at two levels The heaters are simple high power resistors that deliver 1 or 10 Watts at the recommended supply voltage of 12 VDC Switchi...

Page 16: ...the level of 120 W m2 The reference measurement of direct radiation can be done using a radiation sensor with a limited field of view a Pyrheliometer and pointing this towards the sun typically done...

Page 17: ...tectors D1 D2 and D3 The way that CSD 1 works is as follows CSD 1 has three detectors D1 D2 and D3 D1 detects all the solar radiation direct and diffuse D2 and D3 only cover part of the sky the part t...

Page 18: ...by the estimated value of the diffuse radiation to result in the direct radiation Direct radiation D1 C smallest of D2 and D3 Where D1 D2 D3 are the signals of detectors D1 D2 D3 C is a geometry corr...

Page 19: ...tude Required power supply without heating 0 1 W at 12 VDC heating low level 1 W 0 1 W at 12 VDC heating high level 10 W 1 W at 12 VDC 2 1 SPECIFICATIONS OF THE SUNSHINE DURATION MEASUREMENT Estimated...

Page 20: ...HEATER Power supply heating low level 1 W 0 1 W at 12VDC Power supply heating high level 10 W 1 W at 12VDC Expected temperature range for melting ice heating low level only dew removal heating high l...

Page 21: ...cant If too much water is deposited the instrument should be dried inside If the orientation and window are satisfactory check the power supply The voltage input between the connections J15 and J16 J1...

Page 22: ...t sunny not sunny 1 V sunny 0 V not sunny pin 2 output direct signal 1 mV W m2 pin 3 reference voltage 1 115 V 0 003 V pin 4 un buffered direct signal same as pin 2 pin 5 stabilized power supply 5 V 0...

Page 23: ...beam spectrum should resemble the solar spectrum as close as possible and have a small opening angle The procedure for calibration by the customer will generally have an accuracy of 5 The general poli...

Page 24: ...ons The average is called FD3 7 FD1 should equal REF1 If the deviation is more than 5 it is suggested to replace the resistors at location 1 RES1 to adjust the sensitivity The value of the replacement...

Page 25: ...f the resistors is 3 RES3 REP3 total old resistance 3 3 REF1 FD3 10 It is recommended to check the result of the resistor replacement by performing a second calibration check Recommended resistors Typ...

Page 26: ...26 figure 4 1 postion of the calibration resistors and connector J18...

Page 27: ...27 5 DELIVERY Delivery of CSD 1 includes the following items CSD 1 sunshine duration meter 0342901 CSD 1 manual 0342300...

Page 28: ...28...

Page 29: ...TS The following spare parts are available Silica gel 2643943 1 kg Complete drying cartridge consisting of clamp spring 0305165 drying cartridge without cover 9012106 cover for cartridge 9012107 rubbe...

Page 30: ...30...

Page 31: ...31 7 ELECTRICAL DIAGRAMS The following contains electrical diagrams to be used during trouble shooting and adapting the calibration factor Figure 8 1 Electrical diagram CSD1...

Page 32: ...32...

Page 33: ...years This can be done at the Kipp Zonen factory Here recalibration to the highest standards can be performed at low cost Recalibration can usually be performed within four weeks If required urgent r...

Page 34: ...34...

Page 35: ...el Qty Requested delivery time I intend to send the instruments to Kipp Zonen on I would like to receive the instrument s back on Conformation by Kipp Zonen Yes the dates are acceptable to us No unfor...

Page 36: ...36...

Page 37: ...el Qty Requested delivery time I intend to send the instruments to Kipp Zonen on I would like to receive the instrument s back on Conformation by Kipp Zonen Yes the dates are acceptable to us No unfor...

Page 38: ...38...

Page 39: ...el Qty Requested delivery time I intend to send the instruments to Kipp Zonen on I would like to receive the instrument s back on Conformation by Kipp Zonen Yes the dates are acceptable to us No unfor...

Page 40: ...40...

Page 41: ...ce Kipp Zonen S A R L 7 avenue Cl ment Ader ZA Ponroy B t M F 94420 LE PLESSIS TREVISE T 33 1 49 62 4104 F 33 1 49 62 4102 E kipp france kippzonen com Our customer support remains at your disposal for...

Reviews: