MKIV OPERATOR'S MANUAL
120
PMT.WINDOW.TIME
float seconds
Immediate
seconds in a window. Should be an integral
multiple of PMT resolution time.
RESET.TIME.OUT
unsigned integer
seconds
on reset
the length of time allowed for initialization of all
motors. This should be larger than the largest
value of MOTOR.TIME.OUT.
RH.ORIGIN
float volt at 0 RH
immediate
Humidity sensor calibration provided by
manufacturer
RH.SLOPE
Float volt/%
immediate
Humidity sensor calibration provided by
manufacturer
SUPPLY.CONVERSION
[<powerSupply>]
float
Immediate
Number of volts at the output of the supply to
read one volt via the A/D
SUPPLY.DELTA
[<powerSupply>]
float volts
on reset
Permitted (max-nominal and nominal-min)
voltage in watermark recording
SUPPLY.NOMINAL
[powerSupply>]
float volts
on reset
Center voltage in watermark recording
TEMP.DELTA [thermalPoint>] float degrees
on reset
Permitted (max-nominal and nominal-min)
temperature excursion in watermark recording
TEMP.NOMINAL
[<thermalPoint>]
float degrees
on reset
Center temperature in watermark recording
TEMP.ORIGIN
[<thermalPoint>]
float degrees
Immediate
number of degrees Celsius offset for 0 volts at
the A/D
TEMP.SLOPE [<thermalPoint>] Float degrees/volt
immediate
number of degrees Celsius per volt at the A/D
(linear thermistors are used here)
TRACKER.DEBOUNCE.TIME Float seconds
immediate
debounce time for the tracker control switches
USE.B3.FOR.LAMPS
Boolean
immediate
Sets response to command: B,3. If set to YES,
lamps are turned on else the command is
ignored.
VERSION.ELEC
unsigned byte
immediate
Records the electronics version
VOLTS.PER.BIT
Float volts
immediate
Resolution of A/D converter in volts per A/D unit
(bit)
Possible <index>es
<index>
Meaning
possibilities/format
<address>
location within the space
ANSI C integer format
1
<analogPt>
which A/D mux input
integer 0 to 23
<digitalPt>
which digital input
integer 0 to 15
<lamp>
which Brewer internal lamp
0 to 1 or identifier - One of: HG, STD
<motorId>
identifies a motor
see section on I command
<powerSupply>
identifies a power supply
0 to 6 or identifier One of:
HV, +12V, +5V, -12V, +24V, +5VSEC
(secondary supply), -5VSEC
<thermalPoint>
identifies a thermal sensor
0 to 5 or identifier One of:
PMT (photomultiplier), FAN (internal temperature),
BASE, BELOW.SPECTRO,
WINDOW.AREA, EXTERNAL (external temperature)
<trackerSwitch>
identifies a tracker switch
digital input
0 to 4 or identifier One of:
CW, CCW, UP, DOWN, COARSE
Examples
?MOTOR.MAX.ACC[IRIS]
reports the current maximum acceleration for the iris motor.
!PMT.HV.TOLERANCE 50
sets the value of the high voltage tolerance to 50 millivolts.
1
The first character must be a digit. The number interpreted as decimal unless the first digit is a 0 in which
case the second character is tested. If it is 'x' or 'X' the remaining digits are interpreted as a hexadecimal
number. If the second character is a digit the number is interpreted as octal.
Summary of Contents for BREWER MK IV
Page 1: ...INSTRUCTION MANUAL OM BA C230 B Mk IV...
Page 2: ...REVISION HISTORY REV DESCRIPTION DCN DATE APPD B Update 99 16 12 C Update 07 16 10...
Page 3: ...ii...
Page 5: ...2...
Page 7: ...4...
Page 25: ...MKIV OPERATOR S MANUAL 22...
Page 57: ...MKIV OPERATOR S MANUAL 54...
Page 61: ...MKIV OPERATOR S MANUAL 58...
Page 81: ...MKIV OPERATOR S MANUAL 78...
Page 91: ...MKIV OPERATOR S MANUAL 88...
Page 93: ...MKIV OPERATOR S MANUAL 90...
Page 127: ...MKIV OPERATOR S MANUAL 124...
Page 133: ...MKIV OPERATOR S MANUAL 130...
Page 135: ...MKIV OPERATOR S MANUAL 132...