Safety Classification of ME Equipment
Protection against electric shock
Internally powered ME equipment
Applied part
Type BF applied part, including whole
forehead thermometer
Protection against harmful ingress of water
or particulate matter
IP22
Mode of operation
Continuous operation
Note: Not intended to be sterilised. Not for use in oxygen rich environments.
Manufacturer
Zhongshan JinLi Electronic Welghing
EquipmentCO.,LTD.
283
rd
South Min’an Road, Xiaolan Town,
Zhongshan City
,
Guangdong, China
Contact Tel.:+86(760)28133810
Fax:+86(760)28133778
EU-Representative
Wellkang Ltd
Suite B,29 Harley Street
LONDON,W1G 9QR,U.K.
www.CE-Marking.EU
EMC Precaution
•
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information provided
in the ACCOMPANYING DOCUMENTS
•
Wireless communications equipment such as wireless home network devices, mobile
phones, cordless telephones and their base stations, walkie-talkies can a
ff
ect this
equipment and should be kept at least a distance d = 3,3 m away from the
equipment.
(Note. As indicated in Table 6 of IEC 60601-1-2:2015 for ME EQUIPMENT, a typical cell phone
with a maximum output power of 2 W yields d = 3,3 m at an IMMUNITY LEVEL of 3 V/m).