KINLEE FT3010 T-100 User Manual Download Page 1

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Non-Contact Digital Infrared 

Thermometer 

 

FT3010 (T-100) User Manual 

 

Summary of Contents for FT3010 T-100

Page 1: ...Non Contact Digital Infrared Thermometer FT3010 T 100 User Manual ...

Page 2: ...entification 11 Basic Components 11 Display Indicator 12 Operation and Setting Method 12 Installing and Replacing the Batteries 12 Setting Method 12 Advice for First Use 13 Correct Measurement 14 Troubleshooting 16 Care and Maintenance 16 Quality Promised and Service 18 Attachment List 18 Safety Classification of ME Equipment 19 EMC Precaution 19 ...

Page 3: ...ad this user manual Please keep it carefully in case you can refer again someday Declaration of Conformity This device fulfils the provisions of EC directive 93 42 EEC Medical Device Directive This forehead thermometer meets requirements established in ASTM Standard E1965 98 2009 Full responsibility for the conformance of this product to the standard is assumed by Zhongshan Jinli Electronic Weighi...

Page 4: ... details of where and how they can dispose of this item Business users should contact their supplier and check the terms and conditions of the purchase contract This product should not be mixed with other commercial wastes for disposal This product does not contain any hazardous substances Please do not use a combination of old and new batteries Please use the same brand batteries Remove the batte...

Page 5: ...ic shock Disposal of discarded thermometer should be in accordance with national regulations The user should contact the manufacturer if you require further advice and guidance on how to use and maintain this device It is not safe to modify the equipment The materials used on this thermometer have passed the biological performance test They are not potential for an allergic reaction Do not expose ...

Page 6: ... other portable devices that emit electromagnet fields near the unit This may result in inaccurate readings Meanings of Symbols Signs Meaning Type BF applied Part Refer to the instruction manual Manufacturer European community Specific batch code CE mark This marking indicates that the product should not be disposed of with other general household waste ...

Page 7: ...ower supply DC 3V 2 x AA LR6 Transport condition and storage Ambient temperature 20ºC 50ºC Relative humidity RH 15 95 Atmospheric pressure 860 1060 hPa Dimensions 90mm x 43mm x 148mm Weight 100g not including batteries LCD display resolution 0 1ºC 0 1ºF Range of measurements Body temperature 22ºC 43ºC Surface temperature 0ºC 100ºC Measurement accuracy 1 22ºC 42ºC 0 3ºC 2 22ºC 0 5ºC 3 42ºC 0 5ºC Re...

Page 8: ...so be influenced by numerous external factors Normal Temperature According to Age Age Method Normal Temp Range ºC Normal Temp Range ºF 0 2 years 36 4 to 38 0 97 5 to 100 4 3 10 years 36 1 to 37 8 97 0 to 100 0 11 65 years 35 9 to 37 6 96 6 to 99 7 65 years 35 8 to 37 5 96 4 to 99 5 Attention ASTM laboratory accuracy requirements in the display of 37 to 39 98 to 102 for Skin IR thermometers is 0 3 ...

Page 9: ...entification Basic Components Display Installing and Replacing the Batteries 1 Push the battery cover back open the cover from the body 2 Insert two AA batteries as indicated and then replace the battery cover ...

Page 10: ...ential to check the forehead thermometer and change as needed as follows Take the temperature of the same person using a conventional thermometer you will get 37 0 98 6 for instance Take the temperature of the same person using the forehead thermometer keeping the 50 to 150 mm distance between the thermometer and the forehead Take care to remove any obstacle which could alter the measurement hair ...

Page 11: ...teries System fault Speak to the dealer in which you purchased the device Ambient temperature too high Please follow the guidance in the user manual regarding ambient temperature Ambient temperature too low Measuring temperature too high Stop measuring Measuring temperature too low ...

Page 12: ... a soft dry cloth It is regular recommended to do the recalibration every two years to ensure functioning correct Expected Serviced Life 2 Years The typical service life of the new and unused batteries is 10000 measurements for the operation time is 10s WARNING Do not modify this equipment without authorization from manufacturer Manufacturer will provide circuit diagrams component part lists descr...

Page 13: ...ax 86 760 28133778 EU Representative Wellkang Ltd Suite B 29 Harley Street LONDON W1G 9QR U K www CE Marking EU EMC Precaution MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the ACCOMPANYING DOCUMENTS Wireless communications equipment such as wireless home network devices mobile phones...

Page 14: ......

Page 15: ......

Page 16: ......