Quick Reference Guide 

 KI-TK800 Series MPO Visual Cable Verifier

 

Revision 2 

Date: 23 Nov 2017 

 

 

          Page 

2

 of 

4

 

 

KI 28800 series Light Source operation 

 

Only  insert  MPO 

female  (no  pin)  connector  into  instrument’s 

optical adaptor(s). You can use either MM or SM fiber.  

 

To turn instrument on press [On/Off]. 

 

To view firmware version, hold [F4] then press [On/Off].  

 

To view all display segments, press and hold[On/Off], [F1] at the 
same time until it stops beeping. 

 

To defeat auto power-off

,

 hold [On/Off] for 3 seconds at turn on 

until ‘Perm’ is displayed.

 

 

Low battery is indicated with a battery symbol. 

 

To turn on emitters, press [F1/Source]. 

 

To turn off emitter, and return to Home menu, press [Home]. 

 

To turn on back light press [ 

]. 

 

KI 6608 40x magnification microscope operation 

 

If needed, replace universal optical adaptor with an 
MPO12xn or MPO16xn adaptor. 

 

Before inserting any MPO connector, check the scope 
MPO adaptor lateral position, and If necessary, use the 
MPO adaptor alignment screw to adjust the connector to 
the mid-position. 

 

Rotate microscope to the position at which the “TOP” label 

on adaptor is facing up.  Note that image is reversed when 
viewed through microscope. 

  

   

 

Insert connector to be inspected. 

 

Press and hold front switch of blue toggle button to use green 
oblique illumination. This is very well to show dirt on the end face. 

 

While holding eye piece to your eye, adjust focus knob to achieve 
a clear image of connector tip.  

When image is in focus, the centre 

of XY-alignment screw is positioned near the white arrow on the 
right of focus knob

.  

 

Rotate XY-alignment screw to position image of connector tip at 
the centre. 

 

With  “TOP”  label  facing  up

,  determine  MPO  cable  type  by 

comparing the pattern image of microscope with that shown on 

MPO  Cable  Verification  Pattern  Guide

  which  comes  with  the 

instrument.   

 

Releasing the front switch of toggle button now will change to 
microscope illumination to coaxial mode. 

 

To turn off illumination manually, press rear button of blue toggle 
button, or it will also automatically turn off later. 

 

To  disable  auto-off  timer,  power  unit  with  external  supply  via 
USB.