KI 9000 UM-10
5
INTRODUCTION AND APPLICATIONS
The KI 9000A Series Optical Power Meter and Optical Light Source
offer both ease of use and superb measurement confidence. They are
very convenient to use, and have reduced cost of ownership.
This equipment can be effectively used by installers, technicians and
engineers to test all types of fiber optic systems in various applications
including SMF, MMF, POF, PCS and ribbon fiber:
Tx / Rx absolute power levels in dBm
Optical loss in dB
Continuity testing with the versatile test tone features
The interchangeable optical connectors are drop and dust protected
by a snap on cover. A wide variety of connector styles are available.
The instruments feature a tough polycarbonate housing with shock
absorbent sides and corners, which has passed extensive drop
testing.
Calibration can be performed by any suitably equipped laboratory. The
recommended re-calibration cycle is 3 years.
Power Meter features
The standard power meters work with fiber core diameter up to 200
micron, with both PC and APC polish connectors.
Large area detector power meters work with fibers with up to 3 mm
active area including MPO/ MTP connectors with up to 72 fibers.
Detector options include Germanium (Ge), Indium Gallium Arsenide
(InGaAs), Silicon (Si), high power, large area and wavelength
selective.
Power meter calibration options are available from 470 nm to 1625 nm
with specified accuracy at power levels from +24 to - 60 dBm.
The display shows dBm, dB and linear units, and display hold is useful
for data recording.
Superior measurement confidence is achieved with a unique Total
Uncertainty Specification, which covers the full temperature,
measurement and connector range. A calibration certificate is
supplied.
The sensitive optical tone detector displays the actual measured tone
frequency in Hz. If a standard tone is detected, the buzzer sounds,
which is useful for fiber identification and continuity testing. MultiFiber
ID (new to A series) can also separately identify any of 12 test tones
from a compatible source, greatly speeding up continuity / polarity
testing.
Power stability testing is easily using the max/min recording function.
TamperLock mode (new to A series) is a useful way to lock the
instrument features to enforce pre-set & error free operation.
TamperLock operation is auditable and can be defeated on-site if
needed.
Slow Mode (new to A series) can be used to increase signal averaging
for noisy signals.