Fig 4    unit:  lb oz     

     

Fig 5   unit:  oz

3. Working temperature:

 5

ºC

- 35

ºC

4. Power supply:

 12Vdc, 600mA AC adapter, with positive center or 6V4AH rechargeable battery.

Note: 
 

 

. When the rechargeable battery is fully recharged, the scale 

can work for 20 hours continuously.

 

 

. When the power supply is lower than 5.4V(+0.1V), the scale 

       

will be turned off automatically .

 

 

. When the power supply is lower than 5.7V(+0.1V) but over 

      

5.4V(+0.1V), the scale will alternatively display ‘Lo.bAT” and the 

other current display. 

5. Working current:

 < 150mA

6. Recharging current:

 < 400mA

Key function

1. ZERO /ON/OFF:

1.1 When scale is off, press this button to turn the scale ON. The scale will enter self-examination, show the working 

voltage “x.x Udc”, display full capacity “F3.000kg”, “F6.000kg” or “F15.000kg”.  The scale will get  proper power-on-zero-
point and display zero reading. (Note: the proper (power-on-zero-point range is: (calibration zero-point) +20% FS)

1.2 When the scale is on, press this key for less than 5 seconds, and ZERO key function will activate; press this 

button for more than 5 seconds, and ON/OFF key function will activate:

ZERO:

ZERO key range is (power-on zero-point) +5% FS. ZERO function can only be activated when the scale is in 

stable weighing mode and the activation of the ZERO key will also clear the recorded tare weight.  The ZERO LED light 
will be on, NET LED will be off. ZERO function will be not activated when the current weight is not in the range of (power-
on zero-point) +5% FS, and it will display “0------” or “0_____”, then will go back to previous weight reading. 

ON/OFF:

By pressing this key for more than 5 seconds, the scale will display OFF, and turn off after releasing the key. 

2. TARE /UNIT

Press this button for less than 5 seconds, the TARE key function will activate. Press this button for more than 5 seconds, 
UNIT key function will activate:

TARE:

TARE range for this scale is up to 102%FS. This function should be operated when the scale is in stable mode. 

When the TARE function is active, the “NET” LED will be on. [Note: (Gross weight = Net  Tare weight ≤102% 
FS)

UNIT: 

To change the weight unit, hold the TARE/UNIT button for more than 5 seconds or until your desired weight unit 

appears.

Calibration 

Calibration requires technical knowledge and standard test weights.  Calibration should only be performed by a scale technician.  
Check your local phone book for your nearest scale repair center.  If the she scale technician does not have the Kilotech calibration 
procedure and access codes, this can be obtained by calling the Kilotech Service Center at 877-328-5988 (USA) or 800-694-4445 
(Canada).