KPC 2000, Operations Manual E 12202017, Page 4 of 5
ZERO/ON/OFF:
• When the scale is off, press the ZERO /ON/OFF key to turn the scale ON. The scale will enter
self-examination. The working voltage “x.x Udc” will be displayed. The full capacity will be displayed as
“F3.000kg”, “F6.000kg” or “F15.000kg”.
• When the scale is on, press the ZERO /ON/OFF key for less than 5 seconds. The ZERO key function will activate.
• The Zero key can only be activated within 5% of the scale’s capacity.
• The scale should be stable and the platter empty.
• Press the ZERO /ON/OFF key for more than 5 seconds to activate the ON/OFF function
TARE /UNIT:
• Press this button for less than 5 seconds, the TARE key function will activate. Press this button for more than
5 seconds, UNIT key function will activate:
• TARE: TARE range for this scale is up to 102%FS. This function should be operated when the scale is in stable
mode. When the TARE function is active, the “NET” LED will be on. [Note: (Gross weight = Net Tare
weight ≤102% FS)
• UNIT: To change the weight unit, hold the TARE/UNIT button for more than 5 seconds or until your desired
weight unit appears.
CALIBRATION:
• See Calibration video or Calibration Manual
AUTO-OFF FUNCTION:
• When the scale is on, simultaneously press and hold down the TARE and ZERO buttons for more than
5 seconds till the “A.oFF.x” is displayed. X = 0,1…9:
• The AUTO-OFF FUNCTION is used to save battery power.
• You can set the amount of time from 0-9 minutes. If you set to 2 minutes, the scale will shut off after
2 minutes in idle mode.
• When X = 0, the auto-power off function is disabled.
• Press the TARE button to choose the auto-off time and press ZERO button to confirm and save the setting.
LOW BATTERY SIGNAL:
The scale may be powered by the rechargeable battery or 110 AC CUL US approved adapter. When “Lo.bAt” is dis-
played alternatively, this indicates that the battery is low and will require a charging. The voltage of internal work-
ing is below 5.7V(+0.1V)