6-12
TOS9213S/ TOS9213AS
6.3
INTERLOCK Connector
As a means of ensuring the safety of the operator, the tester is equipped with an
interlock function to cut off output in coordination with external devices.
Once the interlock input terminals are opened, the interlock function activates to
shift the tester to the PROTECTION status (“PROTECTION” lights up). The output
is cut off to disable testing. In this status, it is impossible to start a test using the
START switch (
*
1) or cancel the PROTECTION status using the STOP switch (
*
2).
The interlock function restricts the tester’s output from outside to ensure the safety
of the operator.
If the interlock function shifts the tester to the PROTECTION status, short-circuit
the interlock input terminals to disable the INTERLOCK signal. Then, cancel the
status using the STOP switch.
*
1 Includes the REMOTE terminal, START signal for SIGNAL I/O, and START
command for GPIB and RS-232C
*
2 Includes the REMOTE terminal, STOP signal for SIGNAL I/O, and STOP com-
mand for GPIB and RS-232C
•
The first time the tester is t
u
rned on follo
w
ing deli
v
ery, the interlock f
u
nc-
tion acti
v
ates and testing is disa
b
led. An interlock j
u
mper is pro
v
ided
w
ith
the prod
u
ct to
b
e
u
sed for performance checks. Use this j
u
mper only to
temporarily cancel the protection f
u
nction.
W
hen installing the tester,
u
se the interlock f
u
nction as m
u
ch as possi
b
le
to ens
u
re safety.
W
hen
u
sing jigs and de
v
ices in
w
ithstanding
v
oltage and
ins
u
lation resistance testing, pro
v
ide a safety co
v
er that encloses the DUT
and c
u
ts off o
u
tp
u
t
w
hile the co
v
er is open. Alternati
v
ely, to pre
v
ent electric
shock, set
u
p a safety fence aro
u
nd the testing area that c
u
ts off o
u
tp
u
t in
coordination
w
ith door mo
v
ements.
Circuit of the interlock input block
NOTE
WARNING
INTERLOCK
Door open/close switch
Internal circuit of
TOS9213S/ TOS9213AS
+12 V
+
–
1 k
1 k
3.3 k
5.6 V