TOS9200/9201
Basic Operations
3-17
(A) Connecting the 149-10A
(B) Selecting GU
ARD when the DUT is g
rounded
Fig.3-4
Dangerous connection
•
If the LO
W ter
minal of this tester is connected to the HIGH or LO
W ter
mi
-
nal of the ear
th contin
uity tester
T
OS6200, the ammeter will be subject to
measurement errors if GU
ARD is selected, as the resistor inside the
T
OS6200 is connected in par
allel to the tester’
s ammeter
.
Theref
ore
, to
use the
T
OS6200, a
v
oid connecting these ter
minals or select LO
W
.
Select GU
ARD only when DUTs such as small electronic components and jigs are
completely “
fl
oating” electrically
.
The GU
ARD mode enables high-sensiti
vity
, high accurac
y measurements to be
made, as it e
xcludes the current by stray capacity and insulation resistance of the
test leadwire and jigs.
Fig.3-5
Selecting GU
ARD
Setting to GUARD
Ammeter
TOS9200/9201
149-10A
HIGH VOLTAGE terminal
LOW
terminal
D
U
T
Stray capacity,
insulation
resistance of test
lead wire and jigs
Setting to GUARD
Ammeter
TOS9200/9201
HIGH VOLTAGE terminal
LOW
terminal
D
U
T
CAUTION
Stray capacity,
insulation
resistance of test
lead wire and jigs
Ammeter
TOS9200/9201
HIGH VOLTAGE terminal
LOW
terminal
D
U
T
Setting to GUARD
Summary of Contents for TOS9200
Page 22: ...P 10 Preface TOS9200 9201...
Page 116: ...3 78 Basic Operations TOS9200 9201...
Page 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Page 148: ...7 4 Maintenance TOS9200 9201...
Page 162: ...8 14 Specifications TOS9200 9201...
Page 170: ...I 4 Index TOS9200 9201...
Page 171: ......