54
TOS5200
SIGNAL I/O Connector
The SIGNAL I/O connector is the D-sub 25-pin connector on the rear panel.
This connector is used to start and stop tests, recall panel memory entries and test modes,
and monitor the status of the TOS5200.
•
Connector on the rear panel
XM3B-2522 D-sub 25-pin female connector (socket), M2.6
x 0.45 screws; manufactured by OMRON Corporation
•
Compatible connector (plug)
D-sub 25-pin male (with fix screw M2.6)
To avoid malfunctions caused by noise, use shielded D-sub
25-pin connectors and a cable that is 2.5 m or less in length.
For information about how to obtain replacement parts, contact your Kikusui agent or distrib-
utor.
For information about how to use these components, see the OMRON Corporation catalogs.
Wire and tool that are necessary to make the connection
SIGNAL I/O specifications
WARNING
Risk of electric shock. Ensure that all devices are off before you connect or discon-
nect cables between them.
CAUTION
Risk of damage to the internal circuitry.
• Keep the signal wire at least 500 mm away from the high-voltage test lead and the DUT.
• Do not short the output voltage circuit.
• If you use the TOS5200 with incomplete connections, burn outs caused by heating may
occur when the output is turned on.
• Do not touch the contacts or attach insulators to them. Doing so may reduce the quality of
the contacts or cause other problems.
• Do not short the +24 V of pin 24 to the chassis or the circuit common.
NC03403-1
SIGNAL I / O
RS232C
DO NOT REMOVE COVERS. NO OPERATOR SERVICEABLE
PARTS INSIDE.
REFER SERVICING TO QUALIFIED SERVICE PERSONNEL.
UNPLUG THE MAINS SUPPLY CORD BEFORE HANDLING
THE CORD OR LOAD WIRES.
10
SIGNAL I/O connector
OMRON corporation:
XM3B-2522
Wire
Single wire: 0.32 mm (AWG28) to 0.65 mm (AWG22) in diameter
Twisted wire: 0.32 mm
2
(AWG22) to 0.08 mm
2
(AWG28)
Wire stripper
A wire stripper that matches the wires listed above
Input signal
Output signal
Low-active control input
Open collector output
High-level input voltage
11 V to 15 V
Output withstanding voltage 30 Vdc
Low-level input voltage
0 V to 4 V
Output saturation voltage
Approx. 1.1 V
at 25 °C
Low-level input current
5 mA max.
1
1
Excluding the interlock signal
Maximum output current
400 mA total
Input time width
5 ms minimum
Summary of Contents for TOS5200
Page 7: ...TOS5200 7 C Test Start Operation and Display 75 D Troubleshooting 76 Index 79 ...
Page 18: ...18 TOS5200 This page is intentionally blank ...
Page 19: ...Installation and Preparation This chapter describes how to prepare this product for use ...
Page 60: ...60 TOS5200 This page is intentionally blank ...
Page 61: ...Maintenance This chapter explains daily inspections such as measurement checks ...
Page 78: ...78 TOS5200 This page is intentionally blank ...
Page 80: ...80 TOS5200 V version 2 Volt Error Protection 74 W W COMP 32 40 warm up time 64 ...
Page 81: ......