TOS3200
B
System Settings
The system settings consist of the following five screens.
You can enter the system setup screen from the 1/2 screen of each operation mode.
To exit from the system setup screen, press the MANUAL or AUTO key to return to the
screen of the original operation mode.
Setup Items of SYSTEM1/5
Press the SYSTEM key to display SYSTEM1/5.
On SYSTEM1/5, set the items indicated in the following.
SYSTEM1/5
General settings of the TOS3200
SYSTEM2/5
Measurement check
SYSTEM3/5
Management of the single test result data
SYSTEM4/5
Management of the program test result data
SYSTEM5/5
Time settings and calibration management
SYSTEM
MANUAL
I / F
EDIT
AUTO
MEAS
PASS
CONV
MEAS MODE :NORM
BUZ VOL (PASS)
: 3
PASS HOLD : 2.0s
BUZ VOL (FAIL)
: 3
CONV :OFF
CONTRAST
:
5
SELV :OFF LINE
BRK(AUTO)
:ON
SELV
<NEXT>
SYSTEM1 / 5
1
2
3
4
5
6
7
8
9
Item
Description
Panel operation
1
MEAS MODE
Sets whether to hold the maximum value of the measured values.
F1 key
NORM
Normal measurement (not hold the maximum value)
Rotary knob
MAX
Hold the maximum value.
Displays the maximum value during the measurement
period.
2
PASS HOLD
Sets the time to hold the PASS judgment.
F2 key
Time
Selectable range: 0.2 s to 10.0 s
Rotary knob
HOLD
Hold until the STOP switch is pressed.
3
CONV
Displays the converted measured current based on the specified line
voltage.
Available in the TC measurement or PCC Measurement Mode.
F3 key
Voltage
Selectable range: 80.0 V to 300.0 V
Rotary knob
OFF
Disable the conversion display.
Summary of Contents for TOS3200
Page 15: ...General Description This chapter gives an overview and describes the features...
Page 20: ...20 TOS3200 Options This page has been intentionally left blank...
Page 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Page 105: ...Specifications This chapter gives the specifications and external dimensions...
Page 129: ......