116
TOS3200
A
A List of Default Settings
Initializing the TOS3200
While holding down the SHIFT key, turn
on the POWER switch. The TOS3200 dis-
plays the firmware version followed by
“INITIALIZING!” and starts to initialize.
If you initialize the TOS3200, various set-
tings such as the test conditions and the
saved data will be set to the default values
indicated below.
Default values
INITIALIZING!
Item
Command
Setting after
initialization
Setting after sending the
command
*RST
*RCL
Operation mode
FUNC
TC
←
Depends on the
memory contents.
TC
measurement
conditions
(TC)
PROBE
TC:PROB
ENCPE
←
Depends on the
memory contents.
POL
TC:POL
NORM
COND
TC:COND
NORM
LOWER
TC:LIM:LOW
30 µA
LOWER ON/OFF
TC:LIM:LOW:STAT
OFF
UPPER
TC:LIM:UPP
30 mA
UPPER ON/OFF
TC:LIM:UPP:STAT
ON
TIMER
TC:TIM
10 s
TIMER ON/OFF
TC:TIM:STAT
OFF
WAIT
TC:WAIT
1 s
WAIT ON/OFF
TC:WAIT:STAT
OFF
NTWK
TC:NETW
A
MODE
TC:MODE
RMS
RANGE
TC:RANG:SEL
AUTO
PCC
measurement
conditions
(PCC)
POL
PCC:POL
NORM
←
Depends on the
memory contents.
COND
PCC:COND
NORM
LOWER
PCC:LIM:LOW
30 µA
LOWER ON/OFF
PCC:LIM:LOW:STAT
OFF
UPPER
PCC:LIM:UPP
30 mA
UPPER ON/OFF
PCC:LIM:UPP:STAT
ON
TIMER
PCC:TIM
10 s
TIMER ON/OFF
PCC:TIM:STAT
OFF
WAIT
PCC:WAIT
1 s
WAIT ON/OFF
PCC:WAIT:STAT
OFF
MODE
PCC:MODE
RMS
RANGE
PCC:RANGE:SEL
AUTO
Summary of Contents for TOS3200
Page 15: ...General Description This chapter gives an overview and describes the features...
Page 20: ...20 TOS3200 Options This page has been intentionally left blank...
Page 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Page 105: ...Specifications This chapter gives the specifications and external dimensions...
Page 129: ......