External Control

You can use the J1 connector of external control.          

1
2
3
4
5
6
7
8
9

10
11
12
13

14
15
16
17
18
19

20

21
22
23
24
25

1
2
3
4
5
6
7
8
9

10
11
12

14
15
16
17
18
19

20

21
22
23
24
25

13

   

1 VMON

Output voltage monitor.

2 IMON

Output current monitor.

3 A COM

External signal common for pins 1, 2, 4, 
and 14. 

*1

4 EXT-V CV 

CONT

Used to control the output voltage with 

an external voltage

5 A COM

External signal common for pins 1, 2, 4, 
and 14. 

*1

6 EXT-R CV 

CONT

Used to control the output voltage with an external 

resistance.

7 EXT-R CV 

CONT COM

Used to control the output voltage with an ex-
ternal resistance.

8 N.C

Not connected.

9 N.C

Not connected.

10 N.C

Not connected.

11 CV STATUS

On when the PWX is in CV mode.
Photocoupler open collector output. 

*2

12 CC STATUS

On when the PWX is in CC mode.
Photocoupler open collector output. 

*2

13 ALM STATUS On when a protection function has been activated. 

Photocoupler open collector output. 

*2

14 EXT-V CC 

CONT

Used to control the output current with an external 

voltage.

15 A COM

External signal common for pins 1, 2, 4, and 14. 

*1

16 EXT-R CC 

CONT 

Used to control the output current with an external 

resistance.

17 EXT-R CC 

CONT COM

Used to control the output current with an external 

resistance.

18 OUT ON/OFF 

CONT

Output on/off control terminal. 

The output can be turned on and off using exter-

nal contact input.

19 D COM

External signal common for pins 18. 

*1

20 N.C

Not connected.

21 N.C

Not connected.

22 N.C

Not connected.

23 OUT ON 

STATUS

On when the output is on. 
Photocoupler open collector output. 

*2

24 PWR ON 

STATUS

On when the power is on.
Photocoupler open collector output. 

*2

25 STATUS COM Status signals common for pins 11, 12, 13, 23 and 

24.

*1.  During remote sensing, this is the negative (-S) of sensing input. When 

remote sensing is not being performed, this is connected to the negative 
output.

*2.  Open collector output: Max. voltage of 30 V and max. current of 8 mA. 

The status common is floating (isolation voltage or less), it is isolated 
from the control circuit.

Output voltage control

Set the CV control using an external 
voltage or an external resistance in 
the CONFIG settings so that external 
voltage control is enabled (CF05:VoLt/
rES).

• 

Control using an external voltage 

(Vext)

Use pins : 4, 5

• 

Control using an external resis

-

tance (Rext)

Use pins : 6, 7

Output current control

Set the CC control using an external 
voltage or an external resistance in 
the CONFIG settings so that external 
voltage control is enabled (CF04:VoLt/
rES).

• 

Control using an external voltage 

(Vext)

Use pins : 14, 15

• 

Control using an external resis

-

tance (Rext)

Use pins : 16, 17

Output on and off control

To use an external contact to control 
the output on and off states, set the 
appropriate CONFIG parameter (CF06: 
on)

Use pins : 18, 19

External monitoring

• 

External monitoring of the output 

voltage and output current status

Use pins : 1, 2, 3, 5, 15

• 

External monitoring of the

 status

Use pins : 11, 12, 13, 23, 24, 25

For Rext, use a resistor that is rated 

at approx. 10 k

Ω

, 1/2 W or greater, 

that has a low temperature coefficient, 

and that will change little over time. 

Examples of such resistors are metal 

film or wire wound resistors.

CONFIG Settings

Press CONFIG and use the VOLTAGE knob to select the parameter 

that you want to set, and the CURRENT knob to change its value.



Indicates a parameter that is applied immediately.

 

 

r

 Indicates a parameter that is applied when the PMX is turned on.

 

 

¯

Indicates a parameter that is applied when CF34 is executed.

SYSTEM CF00

Resets the panel settings



CF01

Power-on output status

r

CF02

Memory content display



CF03

Key lock



CF04

*1, *2

CC control using an Vext or Rext



CF05

*

1, *2

CV control using an Vext or Rext



CF06

*

1, *2

External control for turning output on and off



CF07

 *2

External control logic for turning output on and off



CF08

Output-on startup state parameter



INTER 

FACE

CF20

Remote interface

r

CF21

SCPI communication error display



LAN

CF30

DHCP

r

 or 

¯

CF31

AUTO IP address

r

 or 

¯

CF32

MANUAL IP address 

r

 or 

¯

CF33

Resets the LAN interface settings (LCI) 



CF34

Restarts the LAN interface (REBOOT) 



CF35

IP address display (1) 

CF36

IP address display (2)

CF37

IP address display (3) 

CF38

IP address display (4)

CF39

MAC address display (1) and (2) 

CF40

MAC address display (3) and (4)

CF41

MAC address display (5) and (6)

USB

CF50

VID (vendor ID) display

CF51

PID (product ID) display

*1.  This parameter affects the PWX when the panel settings are reset (CF00).
*2.  The item can be configured only when the output is off.

Sensing

• Local sensing

This method when the 
load current is small or 
when you do not need to 
consider the load effect 
voltage.

• Remote sensing

This is a feature that sta-
bilizes the output voltage 
across the load by reduc-
ing the influence of volt-
age drops and other ef-
fects caused by the load 
cable resistance.

+S

-S

7mm

STRIP-GAUGE

AWG20-14

Insert the wire 
while holding 
down this 
section with a 
screwdriver.

Terminal Function

  

-

S  

Negative remote 

sensing terminal

 +S

Positive remote 

sensing terminal

The "User's Manual" (included CD-ROM) de-

scribes the following information.

 

Mounting to a Rack

 

 

Connecting to the Output Terminals

 

Options      

Calibration     

 

Troubleshooting

RS232C, USB, and LAN interfaces (LXI) are all 
installed as standard with a Communication fea-
ture.  For information about the remote control, 
see the "Communication interface manual" (in-
cluded CD-ROM). 