PMX35-3A

0-35V 3A

REGUL ATED DC POWER SUPPLY

C

B

A

SHIF T

ALM CLR

F INE

F INE

CURRENT

VOLTAGE

OVP

OCP

SET

LOCK

CONFIG

LOCAL

OUTPUT

CV

CC

A

V

REMOTE

L AN

LOCK

ALARM

PRESET

B

C

A

You can download the most recent manuals 
from the following website.

http://www.kikusui.co.jp/en

PART NO. IB024614 

Feb. 2016

PMX Serie

s

Quick Reference

CURRENT/ FINE

Set the current value, 

change a parameter in the 

CONFIG settings

VOLTAGE/ FINE

Set the voltage value, 

select  the value of a 

CONFIG parameter

OUT PUT

Output on and off

CONFIG/ B

C o n f i g u r e   t h e 

conditions, recall 

and save the preset 

memory B

Status indicators

Voltmeter

Displays the voltage, value 

of a CONFIG parameter, 

cause of an alarm

Ammeter

Displays the current, 

p o w e r ,   C O N F I G 

parameter

SET/ ALM CLR

Set and confirm, release protection 

functions that have been activated

LOCAL/ SHIFT

Switch remote mode, enable the functions 

that are written in blue characters

OCP

OVP/ A

Set and display the protection functions, 

recall and save the preset memory A

LOCK/ C

key lock, recall and save the 

preset memory C

Panel Operations

Using the PWX as a CV or CC Power Supply

When the PWX is in CV mode, the set current is the 
limit of the current that can flow through the load. 
When the PWX is in CC  mode, the set voltage is the 
limit of the voltage that can applied to the load. If the 
limit is reached, the PWX automatically switches the 
CV/ CC mode. 

1

Press SET to change to the setting display.

2

Turn the VOLTAGE knob to set the voltage, 

Turn the CURRENT knob to set the current.

Voltage range: 0 % to 105 % of the rated output 
voltage
Current range: 0 % to 105 % of the rated output 
current

3

Press OUTPUT to turn output on.

The output turns on and off each time you press 
OUTPUT. 

Locking Panel Operations (Key lock)

When the keys are locked, the key and rotary knob 
is invalid. There are three levels of key lock. Use 
CONFIG parameter CF03 to set the level.

 

Loc1: Lock all keys except the OUTPUT and 

memory A, B, and C keys.

 

Loc2: Lock all keys except the OUTPUT key

 

Loc3: Lock all keys and the rotary knob.

Hold down LOCK key until the LOCK LED  

lights. To unlock the keys, hold down LOCK 

key again until the LOCK LED turns off.

To set the PWX to the Factory Default

Hold down LOCK+LOCAL while you turn the 

POWER switch on.

Protection Functions

The PMX is equipped with the following protection functions. When a protection function is activated, an alarm 
occurs, the output is turned off, and the cause of the alarm is indicated.

Press OCP • OVP key for setting the OVP trip point, the OCP trip point.

There are two methods to clear alarm. 

 

Press ALM CLR (SHIFT+SET)

 

Turn the PMX off, fix the problem that caused the alarm, and then turn the PMX on.

Unless you fix th

e problem that caused the alarm to occur, the alarm will occur again.

Overvoltage protection (OVP)

Activated when the output terminal voltage exceeds the set voltage (OVP trip point).   

setting range

 10 % to 110 % of the rated output voltage

Overcurrent protection (OCP)

Activated when  the output current exceeds the set current (OCP trip point).

setting range

 10 % to 110 % of the rated output current

Overheat protection (OHP)

Activated when the internal temperature rises to an abnormal level.

Status indicators

LEDs light when the PMX is in the following states.

CV

Lights during constant voltage mode

   

CC

Lights during constant current mode  

   

ALARM

Lights when a protection function has 

been activated  

   

LOCK

Lights when the keys are locked 

REMOTE Lights during remote control  
LAN

Lights and blinks when the LAN interface 

is in use ( No Fault

/ Fault

 / Stand-

by

 / WEB Identify

 )

PRESET 

A, B, C

Lights when the memory A, B or C values 

are being recalled or saved  

  

Preset Memory Function

You can stores three sets (A, B or C) of the voltage 
and the current.

Saving setting to preset memory

To save a preset memory, press SHIFT+the 

memory key (A, B, or C) to which you want 

to save the settings until their LEDs turn on 

(Hold the keys down). 

Recalling preset memory entries

1

Press SHIFT+the memory key (A, B, or C) 

from which you want to recall the preset 

memory entry.

 If you hold down, the present values in use will 

be saved to the preset memory entry instead.

2

Check the displayed settings, and then 

press SET.

Measured Value Display and Setting Display

• Measured value display

When the SET key LED is off, the present output 
voltage and output current are displayed. You 
can change the output voltage and output cur-
rent while the measured value displayed.

• Setting display

Press SET to light its key and display the present 
output voltage and output current settings.
Press SET again to return to the measured value 
display.

OUT PUT LED

Lights when output is 

turned on