PAD-L
Applied Operation
4-13
•
The release voltage across the control terminals 1 and 2 is approx. 5 V, and the
short circuit current is approx. 1.5 mA.
•
Use external contact points of contact rated 10 VDC min. and 100 mA min.
•
For long-distance wiring, use a small relay and extend the coil side of that
relay.
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
3.
Recheck that the connections.
WARNING
•
After connections are complete, install the terminal cover.
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
NOTE
•
If the offset of the output current is shifted in the positive direction, the output
will not fall to 0 V. If the output cannot be set to 0 V, adjust the I.
OS
variable
resistor. For information on such adjustments, see "Current system adjustment
procedure" described in 6.3.4, "Adjustment procedure", in Chapter 6.
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
C
0
1
2
–
–S
+S
+
3
4
5
6
7
8
9
10
S
S
Relay
Control terminals
Output terminals
2-core shielded wire
or twisted-pair wire
For long-distance
wiring
Extend this line.
Fig. 4-9
Connection for output ON/OFF control (I)