LTP-2

3.

TEST METHOD

1.̲ Connect the probe to the TOS6100 Tester ( or Model 872 Tester)

as follows:

(a) Disconnect the shorting bar from the current/voltage sampling

terminals of the TOS6100 Tester ( or Model 872 Tester)

(b) Connect the DIN plug of the probe-side cable to the remote

control socket of the Tester.

(c) Connect securely the spade lug (larger) of the probe-side cable

to the CURRENT OUTPUT "HIGH" terminal of the Tester,
and  the  other  spade  lug  (smaller)  to  the  VOLTAGE
SAMPLING terminal.

(d) Connect the spade lugs (larger one and smaller one) of the clip-

side cable to the OUTPUT "LOW "terminal of the Tester in the
same manner as above.

2.̲ Apply securely the probe test pin and the clip to test points of the

device to be tested, press the START switch of the probe, and hold
the probe securely in the applied state until the test period preset
on the timer elapses.

3.̲ If the timer is not used, press the STOP switch when a test period

(as required) has elapsed.

4.̲ When the TOS6100 (Model 872) has generated a FAIL signal

(ALARM signal), it can be reset by pressing the STOP switch.

CAUTION

• Do not connect or disconnect the probe test pin and

clip to or from the objective device when the Tester
is delivering its test output. If you do this, the
connecting surfaces of the objective device may be
damaged by sparks of the connect or disconnect
current.

Be sure to connect or disconnect the pin and clip
only when the Tester is not delivering its test output
(when the tester is in the READY (RESET) state).