Test Procedure
•
If the LP02-TOS is used more than 30 A, the contact pin will be
high temperatures. Never touch the contact pin immediately test-
ing is completed. Touching will result in getting burned.
•
Be sure to connect the contact pin and the alligator clip to the
DUT. If the contact pin is removed or the clip is disconnected from
the DUT during testing, result in sparking and potentially damag-
ing the DUT.
•
Configure the required test conditions for the Tester by referring to the opera-
tion manual.
•
To obtain secure identification of FAIL judgment, turning on the FAIL MODE of
the Tester is recommended.
•
Connect the test probe to the Tester according to the “Connecting to the Tes
-
ter” item.
1.
Connect the alligator clip of the LOW test lead to the protective conductor
terminal of the DUT.
2.
Apply the contact pin of the test probe to the test point of the DUT.
3.
Depress the START switch.
This causes the OUTPUT LED to light up; the test current flows from the con-
tact pin.
Do not remove the contact pin from the DUT during testing.
4.
When testing is complete, lift your finger from the START switch.
Do not remove the contact pin from the DUT.
5.
Confirm that the OUTPUT LED is not lit.
6.
Remove the test probe from the DUT.
7.
Disconnect the alligator clip of the LOW test lead from the DUT.
After completing all tests, turn off the POWER switch of the Tester.
WARNING
CAUTION
Testing from the protec-
tive conductor terminal
on the enclosure.
To stop testing, release your
finger from the START switch.
1. To the protective conductor terminal
2. To the test point
3. Depress the START
switch.
The OUTPUT LED
lights up.