2-16
KHA1000
2.6
When the SOURCE Terminal is Connected
Directly to Commercial Power Supply (AC Line)
In directly connecting the SOURCE terminal to the commercial power supply (AC
line) without an AC power supply, such as when measuring rush current, use a
switch or circuit breaker that can disconnect this product from the commercial
power supply (AC line) (Fig.2-9) for safety.
•
You may receive an eclectic shock. To install a switch or circuit
breaker between the SOURCE terminal and commercial power
supply (AC line), be sure to shut off the power supply from the
switchboard by turning off the switches on the switchboard.
•
Set the current rating of the switch greater than the input current of
the EUT.
•
Use a bipolar switch circuit that can simultaneously shut off L and
N.
The L and N polarities of the LOAD and SOURCE terminals conform to IEC
standard measurement category CAT I or CAT II. This category depends on the
input voltage (Table 2-3).
Wire used, and terminal treatment
Table 2-4
Select the wire according to the input current of the EUT. Mount a cramping
terminal to terminate the wire. Use a cramping terminal that has a switch mounting
part with a hole diameter that fits the switch or circuit breaker.
See
WARNING
See
Summary of Contents for KHA1000
Page 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Page 8: ...vi KHA1000 ...
Page 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Page 22: ...xx KHA1000 ...
Page 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Page 30: ...1 8 KHA1000 ...
Page 68: ...3 20 KHA1000 ...
Page 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Page 156: ...5 54 KHA1000 ...
Page 260: ...9 6 KHA1000 ...
Page 261: ...Specifications This chapter explains the specifications of this product 10 ...
Page 272: ...I 4 KHA1000 ...
Page 273: ......