4-10
KFM2151
Remote Control
4.4.3
Default Conditions
Table 4-3 shows the default conditions when *RST, CONF:VOLT,
CONF:VOLT:ALL, or MEAS? is sent and when the KFM2151 is shipped from the
factory. When the power is turned on, the parameters in Table 4-3 are set to the
settings that existed when the POWER switch was turned off.
Ta
b
le 4-3
Defa
u
lt conditions
See
Setting
V
al
u
e
Unit
N
ote
*RST
Factory
Defa
u
lt
CO
N
F
*1
MEAS?
*1.
W
hen CO
N
F:
V
OLT or CO
N
F:
V
OLT:ALL is sent.
CHA
N
:
V
OLT:TERM
*2
*2.
Chan: 1, term±: 01,
v
oltage scanning: on
N
o change
–
Sets the channels on
w
hich to
perform
v
oltage scanning.
CHA
N
:
V
OLT:
N
SEL
1
N
o change
–
Specify the channel on
w
hich
to perform
v
oltage scanning.
CHA
N
:ROUT:TERM
*3
*3.
Chan: 1, term±: 01, impedance scanning: on
N
o change
–
Sets the channels on
w
hich to
perform impedance scanning.
CHA
N
:ROUT:
N
SEL
1
N
o change
–
Specify the channel on
w
hich
to perform impedance
scanning.
V
OLT:RA
N
G
200
*4
*4.
Range specified
b
y CO
N
F:
V
OLT, CO
N
F:
V
OLT:ALL, and MEAS?
–
V
OLT:RA
N
G:AUTO
O
N
–
V
OLT:PROT
200
200
V
V
OLT:PROT:LO
W
-2.00
-2.00
V
Summary of Contents for KFM2151
Page 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Page 4: ...ii KFM2151...
Page 14: ...xii KFM2151 Contents...
Page 35: ...Operation This chapter mainly describes various setup procedures...
Page 54: ...3 20 KFM2151 Operation...
Page 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Page 100: ...4 46 KFM2151 Remote Control...
Page 111: ......