KFM2151
4-9
Remote Control
■
Numeric
A numeric parameter such as a decimal point, option symbol, or measurement unit.
The syntax as a numeric representation is the same as NRf.
MINimum, MAXimum, and DEFault are available as substitutes for declaring
certain values.
Units such as V, A, and W can also be used in a numeric parameter.
If a value that cannot be assigned is entered, the device rounds the value to the
closest possible value.
VOLTage:RANGe 250
Because the range setting is 2 to 200, 200 is returned in response to VOLT:RANG?.
Special form numeric parameters
The special form numeric parameters MINimum, MAXimum, and DEFault can be
used as substitutes for limit values when the parameter is numeric.
In the example below, the voltage protection (UVP) is set to the minimum value.
VOLTage:PROTection:LOWer MINimum
The minimum and maximum values can be inquired for most parameters using
queries.
VOLTage:PROTection:LOWer? MIN
VOLTage:PROTection:LOWer? MAX
Measurement unit
Below are the default measurement units.
•
V (voltage)
The following optional prefixes are supported.
•
M (milli)
• K (kilo)
• U (micro)
•
The unit symbols in the International System of Units contain lowercase
characters. The IEEE standard uses uppercase characters. SCPI commands are
not case sensitive.
•
Commands are accepted even if a measurement unit are not specified.
•
To enter “µ” in the data, use “U” instead.
Summary of Contents for KFM2151
Page 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Page 4: ...ii KFM2151...
Page 14: ...xii KFM2151 Contents...
Page 35: ...Operation This chapter mainly describes various setup procedures...
Page 54: ...3 20 KFM2151 Operation...
Page 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Page 100: ...4 46 KFM2151 Remote Control...
Page 111: ......