KFM2151
2-9
Installation and Preparation
2.5
Connecting the SENSING terminal and the
fuel cell
Use the screwless terminal connector (8 poles) provided. One block contains four
channels of terminals. Connect the SENSING terminal directly to the separator in
order from cell 1.
Fig.2-
8
W
iring the SE
N
SI
N
G terminal
If you want to measure the performance at the electrode terminal, connect to it.
Use AWG24 wires. The recommended amount of covering to be stripped is
approximately 10 mm. Using the strip gauge shown in Fig.2-9 will ensure proper
work.
•
Check the rated temperature of the wires that you are using. Pay close attention
for fuel cells that operate at high temperatures. Even for fuel cells that run at
relatively low temperatures such as a PEFC, the collecting electrode where the
wire is connected may be hot.
Fig.2-9
Connection to the f
u
el cell
1.
Using a scre
w
dri
v
er, insert the
w
ire into the scre
w
less terminal
connector (
8
poles) pro
v
ided as sho
w
n in Fig.2-9.
Fuel cell
SENSING terminal
Cell 1
Cell 2
Cell 3
Wire
1
1
2
2
3
3
4
4
+
-
Fuel cell
Load cable
Twist as much as possible
Twist
Screwless terminal
connector (8 poles)
Wire insertion hole
Insert the wire by pressing with
the tip of a screwdriver.
Return it to the original position
after inserting the wire to fix
the wire in place
Insert into the input
(SENSING) terminal of
the KFM2151
Strip gauge
10 mm
Summary of Contents for KFM2151
Page 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Page 4: ...ii KFM2151...
Page 14: ...xii KFM2151 Contents...
Page 35: ...Operation This chapter mainly describes various setup procedures...
Page 54: ...3 20 KFM2151 Operation...
Page 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Page 100: ...4 46 KFM2151 Remote Control...
Page 111: ......