16
DSI1020/ DSI3020
Installing the Test System
Test system installation procedure
This section explains how to install a test system for “voltage dips, short interruptions and
voltage variations immunity tests.”
Model DSI3020 supports single-phase two-wire, single-phase three-wire, three-phase three-
wire, and three-phase four-wire tests.
Model DSI1020 supports single-phase two-wire tests.
To install the system, you need carry out the following procedure.
Connect the sync signal interface
Connect the test circuit
Connect the communication interface
Note that to use a single-phase three-wire, three-phase three-wire, or three-phase four-wire
output test circuit, you will need to prepare several AC power supplies, attach a single-phase
three-wire output driver or three-phase output driver to each power supply, and connect them.
(This is not necessary if you are using a PCR-LE2 series power supply.)
1
2
3
For details on where to connect the single-phase three-wire or three-phase driver, see
appendix A, “AC Power Supply Connection Diagram” (p. 51)
For details on how to install the driver, see the setup guide of the driver.