71
Remote Control | Message Reference
Trigger system
When triggers are used to make measurements, the product operation varies depending on the continu-
ous measurement function (:INITiate:CONTinuous) and trigger source (:TRIGger:SOURce).
Trigger flow chart
Continuous measurement (:INIT:CONT)
*1
*1.
This can only be set using a command.
ON
OFF
Trigger source
(:TRIG:SOUR)
INT
Performs continuously measurement.
(Trigger flow chart A)
Sending :INITiate or :READ? in the idle
state
*2
applies a trigger. After outputting
the measured value, the product returns
to the idle state.
(Trigger flow chart B)
*2.
Trigger input is not accepted.
EXT
Sending *TRG or a TRIG-IN signal to the
SIGNAL I/O connector or pressing SNGL
TRG on the front panel applies a trigger.
After the measurement is completed, the
product changes to the trigger wait state.
(Trigger flow chart C)
Sending :INITiate or :READ? in the idle
state
causes the product to change to
the trigger wait state. Sending *TRG
*3
or
a TRIG-IN signal to the SIGNAL I/O con-
nector applies a trigger. After outputting
the measured value, the product returns
to the idle state.
(Trigger flow chart D)
*3.
If you send :READ? and the product changes to the trigger wait state, *TRG cannot be used.
Continuous
measurement: ON
Trigger source: INTernal
Measurement output
Measurement
Computation
Delay
A
Continuous
measurement: OFF
Trigger source: INTernal
Measurement output
Measurement
Computation
Delay
B
:INITiate/:READ?/
Idle state
Continuous
measurement: ON
Trigger source: EXTernal
Measurement output
Measurement
Computation
Delay
C
TRIG-IN/*TRG/
SNGL TRG key
Trigger wait
Continuous
measurement: OFF
Trigger source: EXTernal
Measurement output
Measurement
Computation
Delay
D
Trigger wait
Idle state
TRIG-IN/
*TRG
:INITiate
:READ?
TRIG-IN
Summary of Contents for BIM1000 Series