Remote Control | Message Reference
Measurement conditions for the next startup
Saves the present measurement conditions and recalls them at the next startup.
EOM signal output setting
Sets or queries the SIGNAL I/O EOM signal output method.
EOM signal output mode
EOM signal pulse width
Sets the pulse width when :SYSTem:EOM:MODE is set to PULSe.
Sets the current pulse output
When the current pulse output is set to PULSE, measurement current flows only during measurements. By
offsetting the trigger inputs, this function prevents measurement errors caused by mutual interference of
measurement currents, such as when measurements are made using several of this product instruments
simultaneously. This is valid when :INITiate:CONTinuous
is set to OFF or :TRIGger:SOURce
is set to EXTernal.
Program header
:SYSTem:BACKup
Note
Note that there is a limit to the number of times the conditions can be saved (1 million).
Program header
:SYSTem:EOM:MODE
:SYSTem:EOM:MODE?
Value
{HOLD|PULSe}
HOLD:
When :TRIGger:SOURce
is set to INTernal, EOM is turned on after a
measurement is completed until the next measurement starts. When :TRIG-
ger:SOURce is set to EXTernal, EOM is turned on after a measurement is
completed until the next trigger is detected.
PULSe:
Outputs an EOM signal at the pulse width set with :SYSTem:EOM:PULSe.
Response
{HOLD|PULSE}
Program header
:SYSTem:EOM:PULSe
:SYSTem:EOM:PULSe?
Value/Response
<0.001 to 0.099 (NR2)>
Unit
Seconds
Program header
:SYSTem:CURRent
:SYSTem:CURRent?
Value
{CONTinuous|PULSe}
CONTinuous: Runs measurement current continuously.
PULSe:
Runs measurement current only during measurements. If continuous
measurement
is set to on, the operation will be the same as
when CONTinuous is specified.
Response
{CONTINUOUS|PULSE}