background image

Ver.1.1 

 

 

  Lumitester Control Kit 

For Smart/PD-30/PD-20 

Operation manual

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Thank you for purchasing the Lumitester Control Kit. 

To use this kit safely and correctly, 

read this operation manual carefully and follow the instructions. 

 

 

日本語による取扱説明は

 

17

ページからとなります。

 

Summary of Contents for PD-20

Page 1: ...ol Kit For Smart PD 30 PD 20 Operation manual Thank you for purchasing the Lumitester Control Kit To use this kit safely and correctly read this operation manual carefully and follow the instructions 日本語による取扱説明は 17 ページからとなります ...

Page 2: ......

Page 3: ...mponents 9 6 Dimensions and part names 9 7 Operating method 10 8 Judgment of a measurement result 12 9 Maintenance 13 10 Troubleshooting 13 11 Specification of light source positive rod 14 12 Specification of a blank tube negative rod 14 13 Calibration 14 14 Warranty 15 15 Reliability 15 16 Contact 15 ...

Page 4: ...rds the following warning symbols are used in this manual to designate the degree of hazard and damage that might occur if this product is used incorrectly Mandatory This symbol indicates a mandatory action Prohibited This symbol indicates a prohibited action Electrical shock hazard This symbol indicates a risk of electrical shock Fire hazard This symbol indicates a risk of smoke emission or fire ...

Page 5: ...e or repair any part of this product Only our qualified service personnel are allowed to service this unit This unit contains high internal voltages so improper handling may cause electrical shock fire or abnormal operation Do not plug or unplug the AC adaptor with wet or moist hands Doing so may cause electrical shock fire or malfunction When using the AC adaptor always connect it to a power sour...

Page 6: ...e time of charge Pull out the AC DC adaptor from an electric outlet and remove it from a light source when you do not use it Do not soil the glass of a luminescence part If dirt and a crack are given to glass light volume will change Do not cover the hole of this side There are two holes of degassing in the side CAUTION ...

Page 7: ... equipment to liquids reagents and organic solvents That may cause failure discoloration and modification If they are attached wipe off immediately Use it in the range of temperature 10 30ºC and humidity 20 85 In case of Lumitester Smart PD 30 turn off the temperature compensation setting Do not use it at the place where temperature changes violently Do not use it at a place with sunlight or stron...

Page 8: ...where corrosive gases are generated Do not use this unit in locations subject to vibration Do not use this unit in locations exposed to excessive dust or debris Do not leave this unit in unstable or hazardous locations Do not apply excessive shock to this unit and do not drop it Do not place this unit in locations where the temperature is extremely low or near heaters Avoid using this unit in loca...

Page 9: ...7 Disposal of this unit When disposing of this unit comply with the laws and regulations in your area WARNING Do not put unit in fire And do not burn it There is risk of explosion or burst inside it ...

Page 10: ...8 2 List of contents Case 1 Light source 1 Blank tube 1 AC DC adaptor 1 AC power cord A plug type 1 AC power cord C plug type 1 Operation manual 1 Silica gel 1 ...

Page 11: ...hanging is unnecessary with rechargeable Charge time is about 30 seconds Luminescence duration About 15 minutes 5 List of components Name Specification and characteristic Quantity Light source positive rod 1 Blank tube negative rod Typical 0 RLU 1 AC DC adaptor for charge Output DC 5V 3A 1 Power cable for Japan A type 1 Power cable for Europe C type 1 Operation manual Ver 1 1 1 Case for transporta...

Page 12: ...cond or more charge has no problem Additional charge is possible 7 3 Remove a DC plug from a jack for charge And remove an AC power cord from outlet Also remove AC plug of an AC DC adaptor from an AC power socket Cautions Do not give stress to a cord AC DC adaptor AC DC adaptor Hold here and connects with Jack and charge Hold here and remove from Jack DC plug AC power cord DC plug ...

Page 13: ...r increases 2 Holding diagonally Holding horizontally 7 6 Measure a blank tube like 7 4 and 7 5 Caution Hold this unit vertically while measurement If you set diagonally or horizontally an error may increase This mark is in front Caution In case of Lumitester Smart PD 30 turn off the temperature compensation setting See Smart or PD 30 operation manual for setting method ...

Page 14: ...light source is dirty Clean it off Failure of Lumitester or a recalibration may be required Please contact the dealer or us 2 When RLU is over 120 of a label display When a RLU display is 1000 1200 or more There is a possibility that the measurement chamber of Lumitester is dirty Clean it off Failure of Lumitester or a recalibration may be required Please contact the dealer or us 3 When the measur...

Page 15: ...anic solvents etc over the instrument If any liquid should be spilt on the instrument wipe the liquid off immediately dry the instrument When dirt remains Take the same treatment as the above mentioned maintenance 10 Troubleshooting When abnormality is observed remove AC power cord from outlet immediately When abnormality is observed such as malfunction burning smell fuming etc there is a danger o...

Page 16: ...tion of charge About 2 mA Charge time 30 seconds or more Operating time 15 minute after the completion of charge Dimensions Diameter Φ12mm Length 134 7mm Weight About 20 g except AC DC adaptor 12 Specification of a blank tube negative rod Item Contents RLU value Typ 0RLU Tolerance of a RLU value Minimum 0RLU Maximum 10RLU Dimensions Diameter about Φ12mm Length about 149 5mm Weight About 10g 13 Cal...

Page 17: ...nd others This warranty is limited to repair or replacement In the case of damage to the peripheral equipment which occurred as a result of use warranty does not cover 15 Reliability Quality and reliability of our products is very important to us However the breakdown occurs in the product including the semiconductor at a certain probability Therefore pay attention to the following item before you...