7
Test Sequence Application
400
Keysight U8903B User’s Guide
Graph
An example of the graph result menu page is displayed as shown in
.
Figure 7-62
Multitone Analyzer > Settings > Result (Spectrum) menu page
Edit Limits
Press the
Edit Limits
softkey to configure the limits settings.
– Track 1st
If Track 1st trace is enabled, the channel 1 limits settings will be copied to the other channels and the limits settings for
the other channels cannot be edited. Any changes made to the channel 1 limits settings will be reproduced in the other
channels limits settings. Disable the Track 1st trace to set the individual channels limits.
– Lower Limit
Enable or disable the lower limit.
– Upper Limit
Enable or disable the upper limit.
– Lower Limit
Set the lower limit value.
– Upper Limit
Set the upper limit value.
Graph Properties
Press the
Graph Properties
softkey to set the graph properties.
– Title
Edit the bar chart title.
– X-axis
– Auto Scale
Enable or disable the X-axis autoscale.
– Unit
Select the X-axis unit type.
– Left
Set the X-axis left edge value. This setting is only available when
Auto Scale
is disabled.
– Right
Set the X-axis right edge value. This setting is only available when
Auto Scale
is disabled.
Table 7-59
TSA > Project > Test > Measurement > Settings > Result (bar chart) menu description (continued)
Menu
Description
Summary of Contents for U8903B
Page 1: ...Keysight U8903B Performance Audio Analyzer User s Guide ...
Page 16: ...16 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 32: ...32 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 278: ...5 Graph Analysis 278 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 290: ...6 Sweep Function 290 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...