C
Appendix C
162
Keysight U8903B Service Guide
Modify a Cable for the U8903B CMRR test
1
Order the following parts as material:
2
Cut one end of the 8120-1840 BNC cable shown in
.
Figure C-1
8120-1840 BNC cable
3
Locate the male end of the XLR cable, then turn the black sleeve counter-clockwise to
loosen it from the metal body.
Figure C-2
Male end of the XLR cable
4
Retrieve the sleeve and wire hood to uncover the soldering wire.
Figure C-3
Uncover the soldering wire
Table C-1
Parts to order
Part
Quantity
U8903-61301, male XLR to female XLR, 2 m, Black
1 piece
8120-1840, BNC male to straight BNC 48-in
1 piece