1
Calibration Procedures
84
Keysight U3606B Service Guide
9
Verify the resistance gain adjustments using the
DC current gain adjustment procedure
Review the
on page 25 and
on page 77 sections before beginning this procedure.
1
Press
[
]
to enter the DC current gain calibration.
2
The display shows the uncalibrated value and the reference value of the
adjustment item.
3
Configure each adjustment item shown in
.
4
Use
or
to select the adjustment item.
5
Apply the input signal shown in the “Input” column of
Table 1-27
Resistance gain adjustment
Adjustment item
Input resistance
Short
Dual banana plug with copper wire short between the two
input terminals
OPEn
Input terminals open (remove any test leads or shorting
plugs from the input terminals)
10.0000 M
Ω
10 M
Ω
1.00000 M
Ω
1 M
Ω
100.000 k
Ω
100 k
Ω
10.0000 k
Ω
10 k
Ω
1000.00
Ω
1000
Ω
100.000
Ω
100
Ω
I
NOTE
If the zero offset adjustment procedure has been recently performed prior to the
DC current gain calibration procedure, the adjustment item “OPEn” can be
omitted.