Calibration Procedures
1
Keysight U3606B Service Guide
49
CV programming and readback accuracy
This test verifies that the constant voltage programming and readback accuracy
are within published specifications.
1
Turn off the instrument and connect a digital multimeter between the front
panel
(red) and (black) output terminals (see
for constant voltage programming and readback accuracy verification,”
2
Turn on the instrument. Press
[Voltage]
to select the constant voltage mode.
3
Output each value in the order shown below. Ensure that the appropriate
range is selected.
Record the output voltage reading on the digital multimeter. Compare the
measurement results to the appropriate test limits in
CV load effect (load regulation)
This test measures the change in the output voltage resulting from a change in
the output current from full load to no load or vice versa.
1
Turn off the instrument and connect a digital multimeter between the front
panel
(red) and (black) output terminals (see
for constant voltage load and line regulation verification,”
2
Turn on the instrument. Press
[Voltage]
to select the constant voltage mode.
Table 1-12
Constant voltage programming and readback accuracy
verification test
Output voltage
Range (parameter)
Error from nominal 1-year
0 V
S1 (30 V)
±0.005 V
30 V
±0.02 V
0 V
S2 (8 V)
±0.005 V
8 V
±0.009 V
0 V
S2m (1000 mV)
±0.001 V
1000 mV
±0.0005 V
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply Service Guide ...
Page 8: ...8 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 12: ...12 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...