1
Calibration Procedures
42
Keysight U3606B Service Guide
Gain verification test
This test checks the full-scale reading accuracy of the instrument. Verification
checks are performed only for those functions and ranges with unique gain
calibration constants.
DC voltage gain verification test
1
Connect the calibrator to the front panel
V
(red) and
LO
(black) input terminals
(see
Figure 1-2, “Test setup for DC voltage and AC voltage gain verification,”
2
Press
[
]
to select the DC voltage function. The DC annunciator will turn on
according to your selection.
3
Select each range in the order shown below. Provide the indicated input
voltage. Compare the measurement results to the appropriate test limits in
. (Be certain to allow for appropriate source settling when using the
Fluke 5520A.)
Table 1-2
Zero offset verification test (continued)
Input
Function
Range
Error from nominal 1-year
Open
Capacitance
1 nF
±0.008 nF
10 nF
±0.05 nF
100 nF
±0.5 nF
1
μ
F
±5 nF
10
μ
F
±0.05
μ
F
100
μ
F
±0.5
μ
F
1000
μ
F
±5
μ
F
10000
μ
F
±50
μ
F
DC current
10 mA
±0.0015 mA
100 mA
±0.005 mA
1 A
±0.00007 A
3 A
±0.00021 A
V
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply Service Guide ...
Page 8: ...8 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 12: ...12 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...