Calibration Procedures
1
Keysight U3606B Service Guide
39
Self-Tests
A brief power-on self-test occurs automatically whenever you turn on the
instrument. This limited test assures that the instrument is capable to operate.
During the self-test, all display segments and annunciators are turned on before
doinG
and
SELF.t
are shown on the primary and upper secondary displays
respectively.
If the self-test fails, the
Error
annunciator is turned on.
Read the errors using the utility menu (see
Chapter 4 of the U3606B User’s Guide,
“Reading error messages,” starting on page 129
), or use the
SYSTem:ERRor?
query from the remote interface (see the
U3606B Programmer’s Reference
).
A list of the possible self-test errors is given in
Chapter 6 of the U3606B User’s
Guide, “List of Error Messages,” starting on page 181
. If repair is required, contact
the nearest Keysight Service Center.
You can initiate a more complete self-test from the utility menu (see
Chapter 4 of
the U3606B User’s Guide, “Performing a self-test,” starting on page 142
), or by
sending the
*TST?
command to the instrument from a remote interface.
The command may take up to 30 seconds to complete. You may need to set an
appropriate remote interface SCPI query time-out value.
NOTE
– The instrument returns to normal operation if all self-tests pass. The
Error
annunciator is turned on if a failure occurs.
– The
*TST?
command returns a
+0
if all self-tests pass or a
+1
if a failure
occurred.
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply Service Guide ...
Page 8: ...8 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 12: ...12 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...