1
Calibration Procedures
102
Keysight U3606B Service Guide
Calibration Error Codes
The following errors indicate failures that may occur during a calibration.
Table 1-34
List of calibration errors
Error code
Error message
701
Cal security pads short
702
Cal secured
703
Invalid secure code
704
Secure code too long
705
Cal aborted
706
Cal value out of range
707
Cal signal measurement out of range
708
Cal signal frequency out of range
709
Cal source unfinished
710
EEPROM write failure
720
Cal DCV offset out of range
721
Cal DCI offset out of range
722
Cal RES offset out of range
723
Cal CAP offset out of range
726
Cal RES open out of range
742
Cal checksum failed, DCV corrections
743
Cal checksum failed, DCI corrections
744
Cal checksum failed, RES corrections
745
Cal checksum failed, ACV corrections
746
Cal checksum failed, ACI corrections
747
Cal checksum failed, FREQ correction
748
Cal checksum failed, CAP corrections
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply Service Guide ...
Page 8: ...8 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 12: ...12 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight U3606B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...