Calibration Procedures
Adjustments procedures
204
U3606A User’s and Service Guide
4
Use
or
to select the adjustment item.
5
Apply the input signal shown in the “Input” column of
6
Enter the actual applied input (see
“Entering adjustment values” on
7
Press
to start the adjustment. The display “CALib” in the lower
secondary display starts flashing to indicate that the calibration is in
progress.
•
Successful completion of the adjustment is indicated by a short beep
and the primary display briefly showing “PASS”.
•
An adjustment failure is indicated by a long beep, the primary
display showing “FAiL” and a calibration error number appearing in
the upper secondary display. Correct the problem and repeat this
procedure.
8
Repeat
through
for each gain adjustment point shown in
.
9
Verify the resistance gain adjustments using the
verification test” on page 160
.
N O T E
Always complete tests in the same order as shown in
.
Table 6-6
Resistance gain adjustment
Adjustment item
Input resistance
Short
Dual banana plug with copper wire short
between the two input terminals
OPEn
Input terminals open (remove any test leads or
shorting plugs from the input terminals)
10.0000 M
Ω
10 M
Ω
1.00000 M
Ω
1 M
Ω
100.000 k
Ω
100 k
Ω
10.0000 k
Ω
10 k
Ω
O U T
S B Y
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...