System Related Operation
4
Utility Menu Summary
U3606A User’s and Service Guide
121
If the power- on or complete self- test fails, the Error annunciator
illuminates, and an error is stored in the error queue. See
for more information
on resolving the error.
Selecting a dBm reference resistance value
The dBm function is logarithmic, and is based on a calculation of power
delivered to a reference resistance, relative to 1 mW. The following
procedure shows you how to select an appropriate dBm reference
resistance value.
1
To access the utility menu, press
Shift
>
Utility
.
2
The first utility menu item (Error) will be shown in the upper
secondary display.
3
Press
or
until the menu item “db.rEF” is shown in the upper
secondary display. The factory default dBm reference resistance value is
600
Ω
.
4
Press
or
to change the dBm reference resistance value
(0001
Ω
to 9999
Ω
). See
for more
information.
5
After selecting an appropriate dBm reference resistance value, press
Shift
>
Save
to save or press
Shift
>
Exit
to exit the edit mode without
saving.
6
Press
Shift
>
Exit
to exit the utility menu.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...