5
Verification and Performance Tests
Self-test
154
U3606A User’s and Service Guide
Self-test
A brief power- on self- test occurs automatically whenever you turn on the
instrument. This limited test assures that the instrument is capable of
operation.
During the self- test, all display segments and annunciators are illuminated
before “doinG” and “SELF.t” is shown on the primary and upper
secondary display respectively.
If the self- test fails, the Error annunciator illuminates.
Read the errors using the utility menu (see
), or use the
SYSTem:ERRor?
query from the remote interface
(see the
U3606A Programmer’s Reference
). A list of the possible self- test
If repair is required, contact the Keysight Service Center.
You can initiate a more complete self- test from the utility menu (see
“Performing a self- test” on page 120
), or by sending the
*TST?
command
to the instrument from a remote interface.
This command may take up to 30 seconds to complete. You may need to
set an appropriate remote interface SCPI query timeout value.
N O T E
•
The instrument returns to normal operation if all the self-tests pass, or
the Error annunciator illuminates if a failure occurs.
•
The
*TST?
command returns a “
+0
” if all the self-tests pass, or a “
+1
”
if a failure occurred.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...