Calibrator Output Operations
2
Keysight U1401B User’s and Service Guide
49
Table 2-1
Default settings for the autoscan output
Mode
±1.5000 V
±15.000
V
±25.000
mA
Step
Amplitude
Time interval
Amplitude
Time interval
Amplitude
Time interval
1
+
1.5000 V
02 sec
+
15.000 V
02 sec
+
00.000 mA
02 sec
2
+
1.2000 V
02 sec
+
12.000 V
02 sec
+
04.000 mA
02 sec
3
+
0.9000 V
02 sec
+
09.000 V
02 sec
+
08.000 mA
02 sec
4
+
0.6000 V
02 sec
+
06.000 V
02 sec
+
12.000 mA
02 sec
5
+
0.3000 V
02 sec
+
03.000 V
02 sec
+
16.000 mA
02 sec
6
+
0.0000 V
02 sec
+
00.000 V
02 sec
+
20.000 mA
02 sec
7
–0.3000 V
02 sec
–03.000 V
02 sec
+
16.000 mA
02 sec
8
–0.6000 V
02 sec
–06.000 V
02 sec
+
12.000 mA
02 sec
9
–0.9000 V
02 sec
–09.000 V
02 sec
+
08.000 mA
02 sec
10
–1.2000 V
02 sec
–12.000 V
02 sec
+
04.000 mA
02 sec
11
–1.5000 V
02 sec
–15.000 V
02 sec
+
00.000 mA
02 sec
12
+
0.0000 V
00 sec
+
00.000 V
00 sec
+
04.000 mA
00 sec
13
+
0.0000 V
00 sec
+
00.000 V
00 sec
+
08.000 mA
00 sec
14
+
0.0000 V
00 sec
+
00.000 V
00 sec
+
12.000 mA
00 sec
15
–1.5000 V
00 sec
–15.000 V
00 sec
+
16.000 mA
00 sec
16
+
0.0000 V
00 sec
+
00.000 V
00 sec
+
20.000 mA
00 sec