Recharging the batteries
Fuse replacement
Replace any blown fuse in the instrument according to the following procedures:
1
Power down the instrument and disconnect all the test leads. Make sure that
the charging adapter is also removed.
2
Remove the battery cover and the batteries.
3
Loosen the three screws found on the bottom of the case and remove the
bottom cover.
4
Take out the circuit board as shown in
.
5
Gently remove the defective fuse by prying one end of the fuse loose and
sliding it out of the fuse bracket.
6
Replace it with a new fuse of the same size and rating. Make sure the new fuse
is centered in the fuse holder.
External AC
adapter jack
Charging indication
Green: Fully charged
Yellow: Charging
Set slide switch
at the CHARGE
position
NOTE
This manual provides only the fuse replacement procedures, but not the fuse
replacement markings.