Keysight U1401B User’s and Service Guide
Maintenance
General maintenance
Besides the hazards mentioned above, dirt or moisture in the terminals can also
distort the readings. The cleaning procedure is outlined below:
1
Turn the instrument off and remove the test leads.
2
Turn the instrument over and shake out any dirt that may have accumulated in
the terminals.
3
Wipe the case with a damp cloth and mild detergent — do not use abrasives or
solvents containing benzine, benzene, toluene, xylene, acetone or similar
chemicals. Also, do not spray cleaner liquid directly onto the instrument,
because it may seep into the case and cause damage. Wipe the contacts in
each terminal with a clean swab moistened in alcohol.
4
Make sure the instrument has dried completely before using it.
CAUTION
Repairs or services which are not covered in this manual should only be
performed by qualified personnel.
WARNING
Make sure that the terminal connections are correct for a particular
measurement before making any measurement. To avoid damaging the
instrument, do not exceed the rated input limit.
WARNING
To avoid electrical shock or damage to the instrument, prevent water from
getting inside the case.