1
Calibration Procedures
34
Keysight U1270 Series Service Guide
16
DCmA
Turn the rotary switch to
the
position.
30 mA
30 mA
±0.065 mA
±0.065 mA
300 mA
300 mA
±0.65 mA
±0.65 mA
17
ACmA
While the rotary switch is
in the
position, press the
key once.
30 mA
30 mA, 20 Hz
-
±0.295 mA
30 mA, 45 Hz
±0.295 mA
±0.205 mA
30 mA, 65 Hz
±0.295 mA
±0.205 mA
30 mA, 1 kHz
±0.295 mA
±0.295 mA
300 mA
300 mA, 20 Hz
-
±2.95 mA
300 mA, 45 Hz
±2.95 mA
±2.05 mA
300 mA, 65 Hz
±2.95 mA
±2.05 mA
300 mA, 1 kHz
±2.95 mA
±2.95 mA
18
Turn the rotary switch to
the
position.
3 A
3 A
±0.0100 A
±0.0100 A
10 A
10 A
±0.04 A
±0.04 A
[h]
CAUTION:
Connect the calibrator to the multimeter's
A
and
COM
terminals before applying the 3 A and 10 A input.
19
ACA
While the rotary switch is in
the
position, press the
key once.
3 A
3 A, 45 Hz
±0.0325 A
±0.0265 A
3 A, 65 Hz
±0.0325 A
±0.0265 A
3 A, 1 kHz
±0.0325 A
±0.0325 A
10 A
10 A, 45 Hz
±0.125 A
±0.105 A
10 A, 65 Hz
±0.125 A
±0.105 A
10 A, 1 kHz
±0.125 A
±0.125 A
Table 1-2
Performance verification tests (continued)
Step Test function
Range
5520 output
Error from nominal 1 year
U1271A
U1272A/
U1273A/U1273AX
Shift
View
Esc
Shift
View
Esc
Summary of Contents for U1271A
Page 1: ...Keysight U1270 Series Handheld Digital Multimeters Service Guide ...
Page 10: ...10 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...16 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...