4
Service and Maintenance
54
Keysight U1241AN User’s and Service Guide
Troubleshooting
If the instrument fails to operate, check the batteries and test leads, and replace
them if necessary. If the instrument still does not function, check the identification
.
When servicing, use the specified replacement parts only. The
the replacement part numbers.
WARNING
To avoid electrical shock, do not perform any service unless you are qualified
to do so.
Table 4-1
Basic troubleshooting hints
Malfunction
Identification
No LCD display after
power ON
Check the battery polarity and replace batteries if necessary. Ensure that the replaced batteries are new
batteries, it is recommended not to mix old batteries with new batteries.
No beeper tone
Check setup mode and verify if the beeper is set to OFF. Then select the desired driving frequency.
Failed on current
measurement
Check the fuse.
Table 4-2
List of replacement part number
Part number
Description
U1241-44407
U1241AN front housing
U1241-64401
Bottom case and lens assembly
U1241-46401
Battery cover
U1241-44701
Multimeter stand
U1241-60002
LCD display
2110-1400
Fast blow fuse 1000 V, 0.44 A (10 mm x 35 mm)
2110-1402
Fast blow fuse 1000 V, 11 A (10 mm x 35 mm)
Summary of Contents for U1241AN
Page 1: ...Keysight U1241AN Handheld Digital Multimeter User s and Service Guide ...
Page 14: ...14 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...16 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...