:WAVeform Commands
30
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
1025
:WAVeform:SOURce:SUBSource
Command Syntax
:WAVeform:SOURce:SUBSource <subsource>
<subsource> ::= {{SUB0 | RX | MOSI | FAST}
| {SUB1 | TX | MISO | SLOW}}
If the :WAVeform:SOURce is SBUS<n> (serial decode), more than one data set
may be available, and this command lets you choose from the available data sets.
When using UART serial decode, this option lets you get "TX" data. (TX is an alias
for SUB1.) The default, SUB0, specifies "RX" data. (RX is an alias for SUB0.)
When using SPI serial decode, this option lets you get "MISO" data. (MISO is an
alias for SUB1.) The default, SUB0, specifies "MOSI" data. (MOSI is an alias for
SUB0.)
When using SENT serial decode, this option lets you get "SLOW" data. (SLOW is
an alias for SUB1.) The default, SUB0, specifies "FAST" data. (FAST is an alias for
SUB0.)
If the :WAVeform:SOURce is not SBUS, or the :SBUS<n>:MODE is not UART, SPI,
or SENT, the only valid subsource is SUB0.
Query Syntax
:WAVeform:SOURce:SUBSource?
The :WAVeform:SOURce:SUBSource? query returns the current waveform
subsource setting.
Return Format
<subsource><NL>
<subsource> ::= {SUB0 | SUB1}
See Also
•
"Introduction to :WAVeform Commands"
•
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...