DTF (Distance to Fault) Measurements
DTF Measurement Settings
DTF Measurement Settings
DTF Settings Table
You can set and view all of the DTF settings, including some calculated values,
on the DTF Settings table. Learn about the calculated values in
How to make settings on the DTF settings table
— Press
Meas Setup 4
.
— Then
Settings
.
— Press
Next Page
and
Previous Page
to view all settings.
— To change a setting:
—Use the
arrows or rotary knob to highlight a setting.
—Numeric settings can be changed by pressing numbers using the
numeric keypad. Then press Enter or select a suffix if available.
—Other settings require you to press
Edit
, then press a softkey to change
the value.
—When finished changing a value, press
Done Edit
.
— Press
Dock Window
to relocate the Settings table to a position relative to
the trace window. The setting persists through a Preset. Choose from the
following:
—
Full (Default setting)
Only the Settings table is shown on the screen.
The trace window is temporarily not shown.
—
Left
The Settings table is shown to the left of the trace window.
—
Bottom
The Settings table is shown below the trace window.
— When finished changing ALL settings, press
Done
to save your settings.
DTF Measurement (Format)
You can select from 3 different DTF Formats.
— Press
Measure 1
— Then choose from:
—
Distance to Fault (dB)
Faults are displayed on the Y-axis in return loss
format, expressed as a positive number in dB.
—
DTF (VSWR)
Faults are displayed on the Y-axis in SWR. Learn more
about SWR at the FieldFox Supplemental Online Help:
http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.ht
Summary of Contents for N9923A
Page 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Page 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Page 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Page 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Page 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Page 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Page 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Page 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Page 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Page 275: ...Installation Note Xxxxx xxxxx 3 ...