Keysight N9927-90001 User’s Guide 95
NA (Network Analyzer) Mode
NA Mode Settings
—Then enter a positive or negative offset value in dB using the numeric
keypad, the
arrows, or the rotary knob. Press a multiplier key.
Learn about multiplier abbreviations.
— And/or press
Slope
- Offsets the data trace by a value that changes linearly
with frequency. The offset slope begins at 0 Hz. and continues to the stop
frequency. The offset value at the start frequency is then calculated from 0
Hz.
—Then enter a positive or negative offset slope value in dB/GHz using
the numeric keypad, the
arrows, or the rotary knob. Press a
multiplier key. Learn about multiplier abbreviations in
.
Electrical Delay
Electrical delay is a mathematical function that simulates a variable length of
lossless transmission line. Use the electrical delay feature to compensate for
the linear phase shift through a device and view only the deviation from linear
phase of the device.
You can set the electrical delay independently for each measurement trace. To
apply an electrical delay to all measurement traces, use Port Extensions. Learn
Learn how to set Phase formats in
.
How to set Electrical Delay
— Press
Scale / Amptd
— Then
More
— Then
Electrical Delay
— Enter a time value using the numeric keypad, the
arrows, or the rotary
knob.
— Press a multiplier key. Learn about multiplier abbreviations in
.
Electrical Delay can also be set using the
Mkr->Delay
feature. Learn how on
“Marker Functions” on page 459
Phase Offset
Phase offset mathematically adjusts the phase measurement by a specified
amount, up to 360°. Use this feature in the following ways:
Improve the display of a phase measurement.
This is similar to the way you
would change the acquisition in an amplitude measurement. Change the
phase response to center or the response on the screen.
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......