Keysight N9038A MXE Service Guide 683
Tests
Amplitude Accuracy at 50 MHz
7.
Set up the receiver by pressing:
Mode Setup
,
EMC Standard
,
None
Input/Output, More, Freq Ref In
,
External
FREQ
Channel
,
Center Freq
,
50
,
MHz
SPAN X Scale
,
2
,
MHz
AMPTD Y Scale
,
Attenuation, Mech Atten, 10, dB
Sweep/Control
,
Sweep Setup
,
Swp Time Rules
,
SA - Accuracy
Meas Setup
,
Average/Hold Number
,
20
,
Enter
Trace/Detector
,
Trace Average
Single
8.
Perform the following steps for each row listed in
:
a.
Set the signal generator amplitude to the value listed in the
Nominal Source Amplitude column in
b.
Set the receiver input attenuator as indicated in the Attenuation
by pressing
AMTD Y Scale
,
Attenuation
,
Mech Atten
, [Value].
c.
Set the receiver span as listed in the Span column of
pressing
SPAN X Scale
,
Span
, [Value].
d.
Record the signal generator amplitude, as measured by the power
meter, in the Power Meter Amplitude column of
e.
On the receiver, press
Restart
.
f.
Wait for the instrument to finish averaging.
g.
On the receiver press Peak Search.
h.
Record the signal amplitude, as measured by the receiver in the
Measured Amplitude column of
i.
Calculate the signal amplitude accuracy error using the following
equation, and record the results under the Amplitude Accuracy
Error column of
:
Amplitude Accuracy Error = Meas_Amp
−
Power_Meter
Table 22-11
Amplitude Accuracy Results (RF Input 1 - Preamp Off)
Nominal
Source
Amplitude
(dBm)
Attenuation
(dB)
Span
Measured
Amplitude
(dBm)
Meas_Amp
Power Meter
Amplitude
(dBm)
Power_Meter
Amplitude
Accuracy
Error
(dB)
Test Limit
(dB)
−
4
10
2 MHz
±1.13 dB
−
9
10
1 MHz
±1.13 dB
−1
4
10
500 kHz
±1.13 dB
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...