Keysight N9038A MXE Service Guide 313
Reference Assembly
A16 Reference Assembly Troubleshooting
50 MHz Calibrator
1.
Disconnect Cable W43 from J701 as shown in
.
2.
Turn the 50 MHz Calibrator signal on by pressing
Input/Output
,
RF
Calibrator
,
50 MHz
on the instrument.
3.
Verify with a spectrum analyzer that the frequency and amplitude of the
50 MHz Calibrator signal at J701 is equal to that specified for it in
, as shown in
.
Figure 9-5
A16 Reference Assembly 50 MHz Calibrator