134 Keysight N9038A MXE Service Guide
Input Selection & Level Control
Input Selection & Level Control Section Purpose
A23 Limiter Assembly
This assembly is used to protect the rest of the instrument from large transient
signals that are often produced by Line Impedance Stabilization Networks
(LISN) which are used for making conducted emissions measurements.
Control for this assembly comes from the A21 RF Preselector Input assembly.
A23 Limiter Assembly Connections
The connections to and from the A23 Limiter assembly are as listed in
below.
Figure 4-1
A23 Limiter Assembly Connections
Table 4-1
A23 Limiter Assembly Signals
Port
Description
To/From
Cable
J102
IF Input
RF Input J2
W26
J103
RF Output
SW1 Cal Switch P1
W27
J108
JTAG Programming
A21 RF Preselector Input Assembly J602
W1
P2
RF Input LED Control
A1A5J1 & A1A6J1
W6
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...