![Keysight N9032B PXA Manual Download Page 17](http://html.mh-extra.com/html/keysight/n9032b-pxa/n9032b-pxa_manual_1961390017.webp)
Installation Note N9032-90004
17
Option EXM, External Mixing Upgrade Kit
Figure 13
W29 and W30 Properly Aligned
5.
Refer to
. Use a ¼” open-end wrench to prevent the SMA female connector on W29
from twisting. Use a 5/16” torque wrench to torque the nut on W30 to 10 inch-pounds.
Figure 14
Torque Cable W30 onto W29