Installation Note N9032-90004
Option EXM, External Mixing Upgrade Kit
Figure 20
Successful License Installation
Alternate Installation Procedure
The License File can be manually installed over USB or LAN by placing the license file in the
following analyzer folder: C:\Program Files\Agilent\licensing.
Verify the License Installation
1.
Before the licenses will be recognized, the XSA application must be restarted. Press
File
,
Exit
.
An Exit Analyzer dialog box will appear; press
Enter
to confirm the exit.
2.
Double-click on the LaunchXSA icon on the Windows desktop. Wait for the XSA application to
finish starting (the analyzer should be sweeping).
3.
Press
System
,
Show
System
on the analyzer to display a list of all displayed options. You
should see the following option listed:
—
N9032B-EXM External Mixing
Verify Optional Functionality
1.
Press
MODE/MEAS, Spectrum Analyzer
,
Swept SA
,
OK
.
2.
Press
Input/Output
and tap
Select Input
.
3.
Verify that there is a selection for “
External Mixer
” below the “
RF
” selection.