
8
Installation Note N9020-90257
Option BBA, Analog Baseband I/Q Inputs Retrofit Kit
Portable Instrument (Option PRC)
1.
Disconnect the instrument from ac power.
2.
. Using the T-20 driver, remove the four screws
(1)
, two on each side, to
detach the Bail Handle Assembly
(2)
from the front frame.
Figure 2
Bail Handle and Front Bumpers Removal (Option PRC)
3.
Using the T-20 driver, remove the four screws
(6)
(two on each side) that attach the four strap
handle plugs
(5)
from both sides of the instrument.
4.
. Using the T-20 driver, remove the four screws including washers
(1)
that
hold the rear bumpers
(2)
in place.
5.
Pull the instrument cover off towards the rear of the instrument.
Make sure any adapters on the front panel are removed.